International Surgery Journal

Hornok Z et al. Int Surg J. 2026 Dec;13(12):2536-2545

https://www.ijsurgery.com

pISSN 2349-3305 | eISSN 2349-2902

Beyond pneumoperitoneum: alternative and innovative uses of the Veress needle — a systematic review

Zita Hornok1*, Katalin Borbála Potyó2


1Independent Researcher

2Primary Care Health Service, Pápa, Hungary

Received: 16 September 2026

Revised: 17 October 2026

Accepted: 07 November 2026

*Correspondence:

Dr. Zita Hornok,

E-mail: hornok.zita@gmail.com

Copyright: © the author(s), publisher and licensee Medip Academy. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

ABSTRACT

The Veress needle is a cornerstone of minimally invasive surgery. Originally developed to induce artificial pneumothorax for tuberculosis treatment, it is now primarily recognized for its role in creating pneumoperitoneum. The objective of this study was to systematically summarize the innovative, off-label applications of this device that extend beyond its conventional use in establishing pneumoperitoneum. In accordance with PRISMA 2020 guidelines, a systematic search was conducted on PubMed, Scopus and Cochrane Library databases to identify and categorize documented non-conventional applications of the Veress needle. Methodological quality was assessed using the Mixed Methods Appraisal Tool. A total of 54 studies met the inclusion criteria and were included in the qualitative synthesis. The included studies were categorized into four domains: surgical access and technical assistance, thoracic applications and emergency decompression, specialized diagnostic and therapeutic applications, and preclinical and experimental validation. Methodological quality was predominantly moderate (83%), with the majority of reports consisting of case series or case reports. To the best of our knowledge, this is the first systematic review highlighting how this nearly century-old invention has evolved into a versatile, multifunctional tool in modern minimally invasive surgery.

Keywords: Veress needle, Systematic review, Off-label use, Surgical innovation

INTRODUCTION

The Veress needle (VN) was developed in 1933 by János Veress, a Hungarian internist, primarily for the treatment of tuberculosis. Its original indication was to establish an artificial pneumothorax, thereby achieving pulmonary collapse to facilitate healing by resting the infected parenchyma. From the 1960s onward, with the widespread adoption of laparoscopy, its clinical application shifted toward the creation of pneumoperitoneum, eventually becoming the cornerstone for laparoscopic access.1

Owing to its simple yet robust design, global accessibility, and cost-effectiveness, the VN possesses significant innovative potential beyond its conventional role.

Consequently, a growing body of literature reports on leveraging these unique mechanical properties for various off-label, alternative, and minimally invasive applications. The objective of this systematic review is to comprehensively evaluate the innovative and off-label applications of the VN published in the literature to date.

METHODS

Search strategy

This systematic review was conducted in accordance with the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 guidelines.2 The protocol was prospectively registered with PROSPERO (CRD420261404228). Although the original protocol specified searching MEDLINE, PubMed, and Scopus, the search was ultimately performed across PubMed, Scopus, and the Cochrane Library on May 24, 2026, to ensure a more comprehensive identification of clinical trials and evidence.

The search was limited to English-language publications regarding the off-label applications of the VN. The complete search strategy, including all Boolean operators and keywords utilized for each database, is provided in the Supplementary Material (Supplementary Appendix 1).

Eligibility criteria

Studies were included if they reported original data on the off-label use of the VN or devices derived from the VN mechanism for indications beyond conventional laparoscopic pneumoperitoneum creation. Eligible study designs encompassed original English-language research (including clinical trials, observational studies, case reports, and technical notes) involving human patients or animal models. To be selected, studies were required to report at least one of the following outcomes: technical feasibility, procedural success rate, safety, or procedure-related complications.

Conversely, studies were excluded if they focused exclusively on standard pneumoperitoneum creation, failed to explicitly identify the VN as the utilized instrument, or consisted of secondary research lacking original data (e.g., reviews, editorials, letters, and conference abstracts for which the full text was unavailable). Inclusion and exclusion criteria are summarized in Table 1.

Table 1: Eligibility criteria.
Table reproduced from source PDF page 2

Study selection

Initial deduplication was performed using Zotero reference management software (version 9.0.4), followed by manual verification. Two authors (Z.H. and K.B.P.) independently screened the titles and abstracts of all identified records. Subsequently, the full texts of potentially eligible reports were obtained and independently assessed for final inclusion. Discrepancies were resolved through discussion until consensus was reached. The complete selection workflow, including specific reasons for exclusion at the full-text stage, is illustrated in the PRISMA flow diagram (Figure).

Methodological quality assessment

To assess the risk of bias, two reviewers (Z.H. and K.B.P.) independently appraised the methodological quality of the included studies using the Mixed Methods Appraisal Tool (MMAT).3 Following the MMAT guidelines, each study was first categorized by study design and subsequently evaluated against five specific quality criteria focused on methodological rigor, including selection bias, measurement validity, and data integrity. Any discrepancies in scoring were resolved through discussion until consensus was reached.

Data extraction

Relevant data from the included studies were extracted into a standardized data extraction form using Microsoft Excel. To ensure accuracy and consistency, the process was performed by one author (Z.H.) and cross-checked by a second author (K.B.P.). The data points recorded for each study encompassed the population or model used, sample size, the specific indication, and the procedure performed. Additionally, key outcomes and any reported complications or safety data associated with the intervention were documented for final synthesis.

Data synthesis

Due to the observed heterogeneity in study designs, participant populations, and clinical indications, a formal meta-analysis was not feasible. Instead, a narrative synthesis was performed in accordance with the Synthesis Without Meta-analysis (SWiM) reporting guidelines.4 Data were grouped and summarized according to the clinical indications for the off-label use of the VN. Results are presented descriptively and supported by summary tables. Where possible, a quantitative summary of key outcomes such as procedural success and the incidence of complications was provided to offer a structured overview of the evidence across the included literature.

RESULTS

Study selection

The initial electronic database search identified 677 records. After the removal of 51 duplicates, 626 unique records were screened based on their titles and abstracts, leading to the exclusion of 507 articles. The remaining 119 reports were sought for retrieval and underwent full-text assessment for eligibility. During this stage, 65 articles were excluded for the following reasons: 54 focused on standard indications rather than off-label applications, 9 utilized ineligible study designs, and the full text was unavailable for 2 reports. Ultimately, 54 studies met all inclusion criteria and were included in the final synthesis. The complete selection process and the flow of information are illustrated in the PRISMA flow diagram (Figure 1).

Study characteristics

The included studies were published over a 43-year period, spanning from 1983 to 2026. While early publications were sporadic, a steady upward trend in the volume of literature has been observed since the turn of the century. Notably, approximately one-third of the total cohort (35.19%; n=19) were published within the last decade, between 2016 and 2026. The earliest report, dating back to 1983, explored the utility of the VN for thoracocentesis, whereas the most recent publication in 2026 focused on controlled capnothorax induction to facilitate diaphragmatic resection. The geographical distribution of the 54 included studies reflects a widespread international interest in alternative VN applications. These studies originated from 22 unique countries across four continents. The Americas contributed the highest volume of publications (n=21; 38.89%), followed by Europe (n=19; 35.19%), with the USA (n=19; 35.19%) and Italy (n=5; 9.26%) being the most prominent individual contributors. Asia accounted for 24.07% (n=13) of the total cohort, led by research from India (n=4; 7.41%), while only a single study originated from the African continent.

A clear economic divide was observed between clinical and experimental applications. Notably, 27.78% (n=5) of the articles related to unconventional surgical access and technical innovations originated from low- and middle-income countries, whereas the vast majority (n = 18; 94.74%) of articles focusing on experimental studies, drug and substance delivery, and diagnostic applications originated from high-income countries. Across the 54 included studies, off-label interventions or experiments using the VN were performed on a total of 1,846 human patients, 141 animal models, 14 cadaveric models, and 16 ex vivo specimens. Sample sizes were not specified in one article, which focused solely on the technical description and feasibility of a novel procedure.

Regarding study designs, case reports and case series represented the largest portion of the literature (n=28; 51.85%). Observational and cohort studies, encompassing both prospective and retrospective comparative designs, accounted for 20.37% (n=11) of the included studies. High-level clinical and experimental evidence was identified in seven randomized controlled trials (n=7; 12.96%), representing the most robust study methodologies within the cohort. Finally, experimental research accounted for 14.81% (n=8) of the total cohort. Detailed characteristics of the included studies are presented in Supplementary Appendix 2.

Figure 1: PRISMA 2020 flow diagram illustrating the study selection process.
Figure 1: PRISMA 2020 flow diagram illustrating the study selection process.

Methodological quality assessment

Based on the MMAT, the cohort consisted of 36 quantitative descriptive studies (66.67%), 11 quantitative non-randomized studies (20.37%), and 7 quantitative randomized controlled trials (12.96%). The overall quality rating was found to be moderate for the majority of the literature (n=45; 83.33%), while high quality was identified in 8 studies (14.81%) and only 1 study (1.85%) was rated as low quality. On average, the studies met 3.72±0.74 of the 5 quality criteria. The methodological quality scores for each study are summarized in Table 2.

Clinical indications for off-label use

The identified literature revealed a broad spectrum of off-label applications for the VN, ranging from critical emergency interventions to novel diagnostic tools. To provide a structured synthesis, these reports were categorized into four primary functional domains.

Surgical access and technical assistance (n=18)

Within this category, the VN served as a versatile access and auxiliary tool. The device provided safe, non-traditional access to the pouch of Douglas for transvaginal procedures, as well as to the retroperitoneum for renal cyst ablation.5-8 Its role as a conduit was particularly prominent in nephrology for the modified insertion of peritoneal dialysis catheters and in pediatric urology for successful antegrade or retrograde ureteral stenting.9–15

Furthermore, the VN functioned as a mechanical aid, serving as a simple liver retractor in bariatric and gastric procedures thereby obviating the need for additional trocars and as a temporary forceps port in hernia ligation.16-18 The VN was utilized as an auxiliary needlescopic port to facilitate complex suturing during embryonic-natural orifice transluminal endoscopic surgery (E-NOTES).19 Finally, the VN acted as a guide for intragastric port placement or fascial port-site closure.20-22

Table 2: Methodological quality assessment of the included studies.
Table reproduced from source PDF page 4, part 1 of 2
Table reproduced from source PDF page 5, part 2 of 2

Thoracic and emergency interventions (n=11)

A subset of reports highlighted the induction of an artificial pneumothorax to prevent iatrogenic lung injury during transthoracic adrenal biopsies, renal cryoablation, and coil localization of lung nodules, as well as to facilitate diaphragmatic resection.23-26 Similarly, in the absence of significant pleural effusion, the VN was employed to create a safe pleural space to ensure the safety of thoracoscopic procedures.27

In the management of pleural collections, the VN served as a safe tool for bedside thoracentesis and pigtail catheter insertion, even in non-traditional patient positions.28,29 Crucially, the device was identified as an emergency rescue tool for the immediate decompression of life-threatening tension states, including tension pneumoperitoneum following resuscitation and intraoperative capnothorax or capnoperitoneum during complex endoscopic procedures.30-33

Specialized diagnostic and therapeutic applications (n=12)

In this category, the VN was utilized for percutaneous peritoneal lavage in blunt trauma patients and for transfascial intraabdominal pressure monitoring to validate endoscopic insufflation.34-36 Furthermore, the device was successfully employed for the atraumatic aspiration and drainage of various fluid collections, including large ovarian, splenic, and paratubal cysts, as well as hepatic abscesses and periprosthetic seromas in patients with breast implants.37-42

In these procedures, the needle’s safety mechanism was critical for preventing fluid spillage and protecting adjacent anatomic structures or implants from accidental injury. As a delivery system, the VN enabled the precise injection of bulking agents (e.g., Teflon) into the bladder neck to restore continence in pediatric patients.43 Additionally, it provided a safe and cost-effective conduit for the intraperitoneal administration of chemotherapy, offering a reliable single-use alternative to permanent indwelling catheters.44,45

Experimental and preclinical studies (n=13)

Preclinical research utilizing animal and cadaveric models primarily focused on optimizing emergency trauma care. These investigations demonstrated that modified VN-based devices offer superior safety and efficacy for thoracic decompression compared to standard catheters.46-50 The Veress needle was used to establish pleural access for the induction of experimental empyema in an animal training model.51 Furthermore, the VN’s utility was validated for minimally invasive access and assistance across various experimental settings. In cadaveric and animal models, the VN served as a lateral neck retractor to enhance visualization during subxyphoid thyroid lobectomy and as an initial percutaneous tool to reach the pericardial space for epicardial pacemaker lead implantation.52,53

Additionally, preclinical studies validated the VN for confirming the localization of the pouch of Douglas and acting as a conduit for auxiliary instrumentation during NOTES cholecystectomy, thereby obviating the need for standard trocars.54,55 Innovative applications also extended to diagnostic monitoring and device development. The VN was successfully used as an acoustic palpation probe for cartilage characterization and for intra-peritoneal pressure monitoring to verify the seal integrity of gastric closures.56,57 Finally, the core mechanism of the VN has inspired the development of novel needle designs that leverage the spring-loaded safety-feature to minimize tissue trauma during renal puncture.58

Procedural success and safety outcomes

Surgical access and technical assistance

Regarding specialized access, VN-assisted transvaginal hydrolaparoscopy under local anaesthesia was demonstrated to be a feasible outpatient procedure, although rare complications including haematoma and rectal perforation were identified.5-7 For retroperitoneal working space creation, the technique facilitated successful laparoscopic completion without conversion, achieving a lower incidence of subcutaneous emphysema and CO2 leakage.8 In the fields of nephrology and paediatric urology, a 100% technical success rate was reached in peritoneal dialysis catheter insertions, while VN-guided stenting proved significantly more efficient than conventional methods.9-15 Furthermore, the VN served as a highly effective tool for liver retraction; its application obviated the need for additional ports, leading to significantly lower postoperative liver enzyme elevations and reduced pain.16-17 In emergency and reconstructive settings, VN-guided intragastric port placement was successfully performed without VN-related complications.20

Similarly, the VN was effectively utilized as an auxiliary forceps port during laparoscopic inguinal hernia ligation, with no recurrences or complications reported during the follow-up period.18 In complex E-NOTES urological reconstructions, the needle further facilitated reconstructive maneuvers and allowed for the avoidance of additional ports without VN-related adverse events.19 Finally, utilizing the VN for port site closure demonstrated a 100% technical success rate and significant cost-effectiveness, while reducing closure times, postoperative pain, and rates of infection or dehiscence.21-22

Thoracic applications and emergency decompression

The VN yielded high technical success and a superior safety profile across thoracic and emergency applications. In pleural access procedures, success rates ranged from 80.6% to 100%, with comparative evidence indicating significantly lower complication rates than conventional techniques. Notably, VN-guided procedures achieved a 96.7% primary therapeutic success rate in paediatric bedside thoracostomies (STEP-TT) and 100% feasibility in US-guided pleural punctures.28,29 In emergency settings, the VN proved to be a reliable tool for rapid decompression. Technical success was achieved in all cases of emergency abdominal and thoracic decompression.30-32

In specialized surgical settings, the VN enabled precise nodule localization for VATS resections and provided effective lung protection via artificial pneumothorax induction.23-25,27 In advanced oncological cases, VN-assisted laparoscopic liver and diaphragmatic resections were feasible.26 Additionally, the needle was successfully utilized for pleural effusion drainage, achieving 100% resolution.33 Across this group, complications remained minimal, with only rare instances of pneumothorax (1.4%) or procedural error reported.

Specialized diagnostic and therapeutic applications

The VN exhibited remarkable clinical versatility within specialized diagnostic and therapeutic settings. The device facilitated the successful decompression and aspiration of ovarian, splenic and paratubal cysts, as well as liver abscesses, consistently achieving technical success while notably ensuring the complete prevention of intraperitoneal spillage.37-41 In diagnostic applications, such as peritoneal lavage, the VN demonstrated absolute reliability and was significantly faster than open techniques.34,35

Further diagnostic utility was evident in US-guided periprosthetic fluid aspiration and intra-abdominal pressure monitoring during transgastric NOTES, where VN-derived values correlated precisely with endoscopic measurements.42,36 Regarding therapeutic interventions, the VN provided a safe and cost-effective conduit for repeated intraperitoneal chemotherapy delivery, with success rates reaching up to 100%.44,45 These procedures were associated with only minimal morbidity notably a 2.3% minor complication rate and no major injuries. Furthermore, the instrument was effectively utilized for periurethral Teflon injection in paediatric patients with neurogenic bladder, resulting in total continence.43

Experimental and preclinical validation

Preclinical studies confirmed the VN’s robustness and versatility, with success rates generally ranging between 85% and 100%. Technical success was demonstrated in complex procedures, including robotic vNOTES, transsigmoid NOTES and subxiphoid thyroidectomy.54,55,52 Furthermore, modified VN (mVN) reduced procedural trauma during percutaneous renal access procedures compared to standard devices.58 In emergency models, the VN outperformed conventional decompression tools, achieving 100% success in tension pneumothorax models versus 21% with standard needles.46-50

Despite demonstrating superior procedural efficiency and resuscitation efficacy, a performance threshold of 57% was identified in extreme pulseless electrical activity models, where larger-bore instruments proved more effective.53 Finally, the VN proved effective in training models achieving 85% success in empyema induction and diagnostic applications like cartilage grading.51,56 These interventions remained safe across all settings, further underscoring the instrument's mechanical stability and safety. The VN was feasibly and effectively used to monitor intraperitoneal pressure during gastrostomy closure.57

DISCUSSION

Historical timeline and technological evolution

The evolutionary trajectory of the VN underscores a remarkable transition from a single-purpose tool to a versatile surgical accessory, capable of overcoming complex technical challenges in both resource-limited and high-tech environments. Early off-label reports emerged in the 1980s, as the VN began to transcend its primary role as a pneumoperitoneum insufflation device. This shift was partially driven by safety concerns associated with traditional blind abdominal entry, which carries inherent risks of visceral injury and haemorrhage. During this period, it was primarily utilized as a catheter for diagnostic peritoneal lavage, pleural fluid aspiration, and intraperitoneal chemotherapy delivery.34,35,44,45 From the 2000s, driven by the global advancement of minimally invasive surgery, the VN gained prominence as a reliable decompression tool in emergency settings and served as an essential auxiliary instrument during early NOTES procedures.30-32,19,36 Notably, within the past six years, the VN has been successfully integrated into advanced robotic-assisted procedures, advanced oncological resections and high-precision diagnostic modalities, such as acoustic cartilage palpation.26,54,56 Crucially, the traditionally risky blind insertion has been increasingly superseded by visual-guided laparoscopic applications, as well as innovative ultrasound-guided and acoustic emission-based sensing techniques.

Geographical and economic disparities

Beyond its chronological evolution, the clinical application of the VN also exhibits distinct geographical disparities. Specifically, low and middle-income countries effectively leverage the device's simplicity and cost-effectiveness to overcome resource limitations in routine surgical practice. In these settings, the VN is utilized to manage complex tasks that would otherwise require expensive, more-sophisticated instrumentation, including ureteral stenting, aspiration of hepatic abscesses, and port site closure.13-16,41,21,22 In contrast, high-income nations primarily focus on pre-clinical research and high-tech diagnostic and therapeutic innovations, such as osteoarthritic cartilage grading, the advancement of robotic and laparoscopic NOTES, intraperitoneal chemotherapy delivery, and percutaneous epicardial pacing lead placement.56,54,44,45,53

Procedural success and technical efficiency

Across the evaluated domains, off-label applications of the VN demonstrated high procedural success rates. Its spring-loaded mechanism and thin profile provide tactile feedback, enabling precise access during procedures such as transvaginal hydrolaparoscopy without extensive tissue dissection.5-7 Furthermore, the VN has been effectively used for ultrasound-guided aspiration of fluid collections, cysts, and abscesses, ensuring accurate access while potentially minimizing the risk of intraperitoneal spillage.37-41 The VN facilites various surgical procedures and may reduce procedural time in applications such as port-site closure and ureteral stenting in pediatric urology.13-15,22 It can also serve as an auxiliary port to simplify complex suturing during natural orifice transluminal endoscopic surgery (NOTES).19 Several studies have clearly demonstrated that, in patients with blunt abdominal trauma, diagnostic peritoneal lavage performed using a VN is significantly faster than the open technique while maintaining comparable diagnostic efficacy.34,35

Complication rates and safety profile

Conventional blind VN insertion during laparoscopic access has historically been associated with the risk of visceral and vascular injuries. In contrast, the reviewed studies describe a range of off-label applications in which the VN may offer procedural safety advantages. However, differences in clinical settings, techniques, and study designs limit direct comparisons of complication rates across these applications. In thoracic or abdominal emergency procedures, VN-based techniques have been used to achieve controlled decompression with limited tissue trauma.

These techniques may offer an alternative to conventional chest tubes or large-bore needles in selected clinical settings.33,46–50 As a conduit, the VN provides a novel access technique for peritoneal dialysis catheter insertion, significantly reducing the overall complication rate and the risk of catheter migration.9-12 When used as a simple retractor or temporary access port in gastrointestinal and inguinal procedures, the VN may reduce the need for additional trocars and thereby potentially minimize tissue trauma, postoperative pain, and analgesic requirements.16-18

Furthermore, VN-mediated induction of artificial pneumothorax and creation of a retroperitoneal working space facilitate thoracoscopic and laparoscopic procedures by improving access to the target area and reducing the risk of iatrogenic organ injury. These applications include transthoracic adrenal biopsy, laparoscopic liver and diaphragmatic resection, pulmonary nodule localization, and laparoscopic renal surgery.23-26,8

Cost-effectiveness

The reviewed literature suggests that the VN may represent a low-cost, readily available option for selected procedures, potentially reducing the need for specialized equipment. Reported applications include port-site closure, use as a delivery conduit for intraperitoneal chemotherapy, intravesical administration of bulking agents, and as a drainage device for liver abscesses.21,22,41,43-45

Limitations

Several limitations should be considered when interpreting the findings of this review. First, the included studies displayed substantial heterogeneity in terms of study design, patient population, clinical indications, and outcome measures, which impeded any meaningful meta-analysis across most applications. Second, the evidence base was predominantly composed of case reports, case series, and small observational studies, thereby limiting the strength of conclusions regarding comparative efficacy and safety. Finally, another key limitation lies in the methodological quality of the included literature, with 83% of the analysed reports classified as 'moderate' quality. Consequently, there is a clear need for higher-level evidence particularly randomized controlled trials to validate these innovative applications and establish definitive clinical guidelines.

CONCLUSION

Consequently, the VN facilitates various complex procedures even in resource-limited settings or under-equipped facilities and serves as a multifunctional therapeutic and diagnostic tool, as well as an indispensable tool in emergency scenarios. Given the remarkably low complication rates associated with its off-label use, the VN represents a highly promising instrument with significant untapped clinical potential.

Funding: No funding sources

Conflict of interest: None declared

Ethical approval: Not required

REFERENCES

  1. 1. Veress J. A needle for the safe use of pneumoperitoneum. Gastroenterologia. 1961;96:150-152. [German].
  2. 2. Page MJ, McKenzie JE, Bossuyt PM, Boutron I, Hoffmann TC, Mulrow CD, et al. The PRISMA 2020 statement: an updated guideline for reporting systematic reviews. BMJ. 2021;372:n71.
  3. 3. Hong QN, Fàbregues S, Bartlett G, Boardman F, Cargo M, Gramatikov P, et al. The Mixed Methods Appraisal Tool (MMAT) version 2018 for information professionals and researchers. Educ Inf. 2018;34:285-291.
  4. 4. Campbell M, McKenzie JE, Sowden A, Katikireddi SV, Brennan SE, Ellis S, et al. Synthesis without meta-analysis (SWiM) in systematic reviews: reporting guideline. BMJ. 2020;368:l6890.
  5. 5. Giampaolino P, Pellicano M, Tommaselli GA, Bifulco G, Simioli S, Nappi C, et al. In-office transvaginal hydrolaparoscopy: a step-by-step, intraoperative pain evaluation. Arch Gynecol Obstet. 2015;292:1373-1377.
  6. 6. Gordts S, Campo R, Brosens I. Office transvaginal hydrolaparoscopy for early diagnosis of pelvic endometriosis and adhesions. J Am Assoc Gynecol Laparosc. 2000;7:45-49.
  7. 7. Gordts S, Campo R, Rombauts L, Brosens I. Transvaginal salpingoscopy: an office procedure for infertility investigation. Fertil Steril. 1998;70:523-526.
  8. 8. Zhang CJ, Chen RF, Zhu HT, Wen RM, Xue S, Li W, et al. Closed technique for the creation of retroperitoneal working space for symptomatic renal cysts. Minim Invasive Ther Allied Technol. 2013;22:110-115.
  9. 9. Asif A, Tawakol J, Khan T, Vieira CF, Byers P, Gadalean F, et al. Modification of the peritoneoscopic technique of peritoneal dialysis catheter insertion: experience of an interventional nephrology program. Semin Dial. 2004;17:171-173.
  10. 10. Jaroszyński A, Miszczuk J, Jadach M, Głuszek S, Dąbrowski W. A new, safe, and effective technique for percutaneous insertion of a peritoneal dialysis catheter. J Clin Med. 2024;13(9):2618.
  11. 11. Ma Y, Liu S, Yang M, Zou Y, Xue D, Liu Y, et al. Association between different peritoneal dialysis catheter placement methods and short-term postoperative complications. BMC Nephrol. 2021;22(1):151.
  12. 12. Napoli M, Russo F, Mastrangelo F. Placement of peritoneal dialysis catheter by percutaneous method with the Veress needle. Adv Perit Dial. 2000;16:165-9.
  13. 13. Babu R, Arora A, Raj N. Stenting antegrade via Veress needle during laparoscopic pyeloplasty (SAVVY technique). J Indian Assoc Pediatr Surg. 2019;24:117-9.
  14. 14. Bishmanov R, Alchinbayev M. Simplified stenting methods during laparoscopic pyeloplasty in children. Electron J Gen Med. 2020;17(5):em236.
  15. 15. Chandrasekharam VVS, Khyati Kiran J, Uday Bhaskar MNS, Phani A. Novel use of the Veress needle for percutaneous retrograde ureteral access after Cohen reimplantation in children. J Pediatr Urol. 2025;21:1947-9.
  16. 16. Ertekin SC, Akbulut G, Turgut E, Akyol H, Ergenç M, Yeğen C. Evaluation of Veress needle as a liver retraction technique in laparoscopic sleeve gastrectomy. Surg Innov. 2025;32:85-93.
  17. 17. Gianni S, De Luca M, Oscar B, Andrea C, Luca B, David A, et al. Veress needle: a simple liver retraction technique for lap band positioning in single-incision laparoscopic surgery (SILS). Obes Surg. 2012;22(1):190-1.
  18. 18. Tatekawa Y. Laparoscopic extracorporeal ligation of hernia defects using an epidural needle and preperitoneal hydrodissection. J Endourol. 2012;26:474-477.
  19. 19. Desai MM, Stein R, Rao P, Canes D, Aron M, Rao PP, et al. Embryonic natural orifice transumbilical endoscopic surgery (E-NOTES) for advanced reconstruction: initial experience. Urology. 2009;73:182-7.
  20. 20. Leeds IL, Barbon C, Beatty CA, Wolfgang CL. Surgical intervention for a foreign body trapped in the stomach. Am Surg. 2015;81(3):E115-6.
  21. 21. Kotakala BK, Mishra RK. Veress needle for port-site closure. World J Lap Surg. 2015;8:39-42.
  22. 22. Ranjani CM, Anish P. A comparative study between hand suture and Veress needle technique closure of port sites in laparoscopic surgeries. Int J Acad Med Pharm. 2026;8:761-765.
  23. 23. Favelier S, Guiu S, Cherblanc V, Cercueil JP, Krausé D, Guiu B. Transthoracic adrenal biopsy procedure using artificial carbon dioxide pneumothorax as outpatient procedure. Cardiovasc Intervent Radiol. 2013;36:1184-1187.
  24. 24. Maeda H, Kajiwara K, Yoshimatsu R, Yamanishi T, Minamiguchi H, Karashima T, et al. Artificially induced pneumothorax with a Veress needle for cryoablation of renal cell carcinoma. Minim Invasive Ther Allied Technol. 2022;31(3):483-6.
  25. 25. Iqbal SI, Molgaard C, Williamson C, Flacke S. Purposeful creation of a pneumothorax and chest tube placement to facilitate CT-guided coil localization of lung nodules before video-assisted thoracoscopic surgical wedge resection. J Vasc Interv Radiol. 2014;25:1133-1138.
  26. 26. Sabatella L, Aliseda D, Martí-Cruchaga P, Zozaya G, López-Olaondo L, Blanco Asensio N, et al. Controlled capnothorax creation for stapled diaphragm resection: a safe maneuver for en-bloc liver and phrenic resection without diaphragm opening. Surg Endosc. 2026;40:4484-4490.
  27. 27. Huan NC, Kho SS, Nyanti LE, Ramarmuty HY, Rahim MAA, Ho RL, et al. Dry medical thoracoscopy with artificial pneumothorax induction using Veress needle. Tuberc Respir Dis (Seoul). 2025;88:181-189.
  28. 28. Hendrickson RJ, Rothenberg SS, Partrick DA, Bensard DD. Thoracostomy in children utilizing a minimally invasive technique. Pediatr Endosurg Innov Tech. 2004;8(3):241-4.
  29. 29. Soldati G, Smargiassi A, Inchingolo R, Sher S, Valente S, Corbo GM. Ultrasound-guided pleural puncture in supine or recumbent lateral position: feasibility study. Multidiscip Respir Med. 2013;8(1):18.
  30. 30. Johnson S, McCracken J, Baidoun F. Tension pneumoperitoneum after bystander cardiopulmonary resuscitation: a case report. Int J Surg Case Rep. 2018;42:227-232.
  31. 31. Jenkins DW Jr, McKinney MK, Szpak MW, Booker JL Jr. Veress needle in the pleural space. South Med J. 1983;76:1383-1385.
  32. 32. Kurian AA, Dunst CM, Sharata A, Bhayani NH, Reavis KM, Swanström LL. Peroral endoscopic esophageal myotomy: defining the learning curve. Gastrointest Endosc. 2013;77(5):719-25.
  33. 33. Rathinam S, Grobler S, Bleetman A, Kink T, Steyn R. Evolved design makes ThoraQuik safe and user friendly in the management of pneumothorax and pleural effusion. Emerg Med J. 2014;31:59-64.
  34. 34. Lockhart CM, Gerding RL, Imbembo AL, Shuck JM. Percutaneous peritoneal lavage using the Veress needle: a preliminary report. J Trauma. 1987;27:1181-1185.
  35. 35. Saunders CJ, Battistella FD, Whetzel TP, Stokes RB. Percutaneous diagnostic peritoneal lavage using a Veress needle versus an open technique: a prospective randomized trial. J Trauma. 1998;44:883-888.
  36. 36. Nau P, Anderson J, Needleman B, Ellison EC, Melvin WS, Hazey JW. Endoscopic peritoneal access and insufflation: natural orifice transluminal endoscopic surgery. Gastrointest Endosc. 2010;71(3):485-9.
  37. 37. Ehrlich PF, Teitelbaum DH, Hirschl RB, Rescorla F. Excision of large cystic ovarian tumors: combining minimally invasive surgery techniques and cancer surgery—the best of both worlds. J Pediatr Surg. 2007;42(5):890-3.
  38. 38. Kilincaslan H, Cipe G, Aydogdu I, Sarac F, Toprak H, Ari E. Pure laparoscopic management of a giant ovarian cyst in an adolescent. Am J Case Rep. 2014;15:4-6.
  39. 39. Seshadri PA, Poulin EC, Mamazza J, Schlachta CM. Needlescopic decapsulation of a splenic epithelial cyst. Can J Surg. 2000;43:303-5.
  40. 40. Bhansakarya R, Subedi S. Laparoscopic management of large right paratubal cyst: a case report. JNMA J Nepal Med Assoc. 2020;58(227):501-4.
  41. 41. Ekwunife CN, Amadi EC. Laparoscopy-guided liver abscess drainage: an additional use of Veress needle. J Surg Tech Case Rep. 2012;4(1):36-8.
  42. 42. Mazzocchi M, Dessy LA, Marchetti F, Marchetti F, Carlesimo B. The use of the Veress needle to drain mammary periprosthetic fluid. In Vivo. 2010;24(2):219-22.
  43. 43. Lund L, Yeung CK. Periurethral injection therapy for urinary incontinence using a laparoscopic port. J Endourol. 2003;17(7):493-5.
  44. 44. Menczer J. Use of the Veress needle for instillation of intraperitoneal chemotherapy. Gynecol Oncol. 1995;59(2):249-50.
  45. 45. Pfeiffer P, Asmussen L, Kvist-Poulsen H, Bertelsen K. Intraperitoneal chemotherapy: introduction of a new single-use delivery system: a preliminary report. Gynecol Oncol. 1989;35:47-49.
  46. 46. Holcomb JB, McManus JG, Kerr ST, Pusateri AE. Needle versus tube thoracostomy in a swine model of traumatic tension hemopneumothorax. Prehosp Emerg Care. 2009;13(1):18-27.
  47. 47. Leatherman ML, Held JM, Fluke LM, McEvoy CS, Inaba K, Grabo D, et al. Relative device stability of anterior versus axillary needle decompression for tension pneumothorax during casualty movement: preliminary analysis of a human cadaver model. J Trauma Acute Care Surg. 2017;83(1 Suppl 1):S136-41.
  48. 48. Leatherman ML, Fluke LM, McEvoy CS, Pokorny DJ, Ricca RL, Martin M, et al. Bigger is better: comparison of alternative devices for tension hemopneumothorax and pulseless electrical activity in a Yorkshire swine model. J Trauma Acute Care Surg. 2017;83(6):1187-94.
  49. 49. Lubin D, Tang AL, Friese RS, Martin M, Green DJ, Jones T, et al. Modified Veress needle decompression of tension pneumothorax: a randomized crossover animal study. J Trauma Acute Care Surg. 2013;75(6):1071-5.
  50. 50. McEvoy CS, Leatherman ML, Held JM, Fluke LM, Ricca RL, Polk T. Larger-caliber alternative devices for decompression of tension hemopneumothorax in the setting of hemorrhagic shock. J Spec Oper Med. 2018;18(4):18-23.
  51. 51. Marecos MC, Torres RA, Bailez MM, Vagni RL, Klappenbach RF. Pediatric thoracoscopic training in an experimental pleural empyema rabbit model. J Laparoendosc Adv Surg Tech A. 2006;16(4):397-9.
  52. 52. Sharata A, Aliabadi-Wahle S, Bhayani NH, Kurian AA, Reavis KM, Dunst CM, et al. Subxyphoid thyroidectomy: a feasibility study. Surg Innov. 2014;21(2):194-7.
  53. 53. Kumthekar RN, Opfermann JD, Mass P, Contento JM, Berul CI. Percutaneous epicardial pacing in infants using direct visualization: a feasibility animal study. J Cardiovasc Electrophysiol. 2023;34(6):1452-8.
  54. 54. Alshiek J, Marroquin J, Shobeiri SA. Vaginal ultrasound-guided pouch of Douglas robotic entry in a live ovine model and human female cadaveric specimens. J Robot Surg. 2022;16(1):73-9.
  55. 55. Meining A, Wilhelm D, Burian M, Dundoulakis M, Schneider A, von Delius S, et al. Development, standardization, and evaluation of NOTES cholecystectomy using a transsigmoid approach in the porcine model: an acute feasibility study. Endoscopy. 2007;39:860-4.
  56. 56. Sühn T, Pandey A, Friebe M, Illanes A, Boese A, Lohmann C. Acoustic sensing of tissue-tool interactions: potential applications in arthroscopic surgery. Curr Dir Biomed Eng. 2020;6:595-8.
  57. 57. Guarner-Argente C, Córdova H, Martínez-Pallí G, Navarro-Ripoll R, Rodríguez-d'Jesús A, de Miguel CR, et al. Gastrotomy closure with a new tissue anchoring device: a porcine survival study. World J Gastroenterol. 2011;17(13):1732-8.
  58. 58. Morozov A, Kalinin N, Androsov A, Gazimiev A, Lerner Y, Gomez Rivas J, et al. A novel and less traumatic needle for kidney puncture: development and preclinical study results. J Endourol. 2023;37(1):93-8.