Original Research Article
Comparison of various methods of temperature measurement in children admitted in the pediatric ward of a tertiary care center
1Department of Pediatrics, Goverment Medical College, Omandurar Goverment Estate, Chennai, Tamilnadu, India
2Department of Pediatrics, Government Stanley Medical College Chennai, Tamilnadu, India
3Department of Pediatrics, Chengalpatu Medical College, Tamilnadu, India
Received: 19 April 2017
Accepted: 18 May 2017
*Correspondence:
Dr. Jayakumar Muthaiyan,
E-mail: mjkumar67@rediffmail.com
Copyright: © the author(s), publisher and licensee Medip Academy. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
ABSTRACT
Background: Temperature measurement is an important procedure in the diagnosis and management of childhood illness. Rectal measurement of temperature is often done with resentment from the child, the parents as well as the physician. Intra aural (Tympanic), temporal artery and axillary measurements are commonly undertaken for temperature recording in children. The aim of the study was to compare the various methods of temperature measurement (intra aural, temporal artery and axillary measurements) in children. This was a hospital based descriptive study (pilot study) from a pediatric tertiary care center.
Methods: This was a hospital based descriptive study (pilot study) from a pediatric tertiary care center. 250 children admitted with history of fever in the pediatric ward. All consecutive children with fever were recruited for the study. Children with ear pain, ear discharge and wax in the ear were excluded.
Results: Inter-group comparison was performed using student paired t-test. There was no difference in the temperature measured by axillary and temporal artery touch when compared with intra aural. Temperature measured by infra-red temporal at 3 cm showed significant lower temperature. Temporal artery and axillary measurements had a correlation coefficient of 0.74 and 0.64 respectively when compared with intra aural measurement.
Conclusions: Temporal artery and axillary measurements are not significantly different from intra aural measurement in children. For the ease of measurement, better correlation, better compliance of the child, temporal artery method can be a reliable preferred temperature measurement in children with fever.
Keywords: Axillary, Temperature, Temporal artery
INTRODUCTION
Temperature measurement is a common procedure undertaken in pediatrics which helps management of febrile illness. The method of measurement varies with the physician and the age of the child.1,2 Measurement of core temperature is the ideal method of assessment of fever in children. Rectal measurement is known to reflect the core temperature. However, the rectal method is neither preferred by the physician nor the child since it is invasive and creates discomfort to the child as well. Though rare, the reported complication of rectal perforation has made them less preferred method of temperature measurement in clinical settings. However recent reports have identified noninvasive infra-red intra aural measurement (tympanic) to be as accurate as rectal temperature.3 Infrared temporal artery and axillary measurement is commonly used in children. This study was undertaken to compare these commonly used methods of temperature measurement in children.
The aim of the study was to compare the infrared intra aural(tympanic), digital axillary and infrared temporal artery temperature measurement among children with fever in the pediatric ward.
METHODS
To compare the infrared intra aural (tympanic), digital axillary and infrared temporal artery temperature measurement among children with fever in the pediatric ward.
This was a hospital based descriptive study. The study was set at the Pediatric department of a semi urban tertiary care Government Institute. The study duration was from June 2016 to September 2016. All children admitted with fever during the study period were chosen for the study. All consecutive children with fever were recruited.
Exclusion criteria
Children with ear infections, ear pain and wax in the ear and neonates (age less than 28 days) were excluded.
Children admitted with history of fever during the study period were chosen for the study. Following informed written consent (with assent in older children) the study population was recruited consecutively. History regarding ear infections or pain in the ear were noted and children were subjected for ear examination. Children with wax in the ear, discharge or perforation in the ear drum were excluded. All children underwent sequential measurement of all the three methods of temperature measurement. Digital axillary, infra-red intra aural and infra-red temporal touching the skin and at 3 cm distance were the methods of temperature measurement. Axillary temperature was measured using the digital thermometer and the temperature was taken as soon as the beep sound was heard with the digital thermometer kept at the axilla. This was followed by temporal method of measurement using the infra-red thermometer kept on the skin pressed across the forehead skin sliding laterally to the hair line in contact with the skin.
The thermometer was held near the temporal region over the skin and then at a distance of 3 cm. The reading was noted for both the measurements from the digital display. This was followed by tympanic measurement using the infra-red thermometer. Disposable aural caps were used for each measurement. The ear canal was pulled downward and backward and the aural probe was inserted in the canal until it snugly got fitted in the ear canal and the digital display of temperature was noted. Children were consequently recruited based on their willingness and were not included more than once in this study. Infra-red aural measurement was compared with emporal and axillary measurement. Mean and standard deviations were calculated. The temperature measured by various methods were compared using the Student t-test. A P-value less than 0.05 was considered statistically significant. Graph pad software version 2016 was used for calculating student t-test. Correlation coefficient was arrived for the various methods of measurement.
RESULTS
The study group comprised of 250 children. Male female distribution revealed male female ratio to be 1.17:1. Among the study group 38 were less than 3 years (15.2%), 99 between 3-7 years (39.6%), 88 between 7-10 years (35.2%) and 25 between 10-12years (10%). Comparison was done between intra aural and the axillary and temporal (touch and 3cm).
The temperature ranged from 97˚F-103.5˚F in tympanic, 97.3˚F-104.1˚F in temporal touch, 97.5˚F-103.5˚F in temporal 3 cm, 95.5˚F-103.5˚F in axillary measurement. The mean and standard deviation of the various methods of measurement has been summarized in table 1.
Various temperature measurements have been compared and shown as scatter plots with correlation coefficient (r) in figure 1-6.
Inter group comparison was done for the intra aural with the other three methods of measurements. results of the comparison are shown in Table 2.
For all the 3 methods (intra-aural, Infra-red touch and axillary), the in-group variability was noted to be very small as denoted by the SD, indicating the good precision in temperature reading of all 3 methods.
Inter-group comparison was performed using student paired t -test it is noted that there was no difference in the temperature measured by axillary and temporal artery touch. Temperature measured by infra-red temporal at 3 cm showed significant lower temperature.
This difference in temperature measurement at 3 cm distance was statistically significant as compared with axillary, intra-aural, infra-red touch. There were no complications of any sort in the study group.
DISCUSSION
Temperature measurement is an essential component of medical diagnosis and planning investigation and treatment in children.1 The best site for measurement of core temperature is the hypothalamus. since it is inaccessible, closest values would be reflected by the pulmonary artery, distal esophagus, bladder and rectum.2 Rectal temperature measurement is the gold standard but is not the preferred mode of measurement both for the physician and the children. It carries the theoretical drawbacks like rare rectal perforation in neonates, spread of infections and slow to reflect the temperature changes.2 Temperature measurement was conventionally done using mercury thermometers in the past. With increasing awareness about environment friendly devices, digital measurement of temperature came into routine use. However, the site of measurement does make a difference in reflecting the core temperature.
Other measurements including oral and axillary measurement is again not the comfortable method of measurement in children. Both procedures need the children to co-operate and also need thorough cleansing after every use. Commonly used methods are the infra-red, intra aural and temporal measurements. methods of measurements Temporal measurement is more comfortable for the child as well as the physician. Aural measurement though close to core temperature needs careful measurement and needs the thermometer caps for individual use. Intra aural measurement, which measures the radiant heat from the tympanic membrane is an alternative to rectal temperature.3-10 This is a safe procedure and does not carry the risk of perforation when introduced in the ear canal. This is acceptable by children. But tympanic measurements will be significantly affected by ear infections and presence of wax in the ears.11
Studies have shown that axillary measurement to be not so accurate with rectal in comparison to temporal artery measurements. Studies have also shown that there is a good correlation between rectal and axillary temperature but the accuracy of axillary temperature is better at higher temperatures than at lower temperatures.12,13 Though different formulas exist to calculate the expected rectal temperature from the measured axillary temperature none can be recommended at present.14,15Comparison of axillary, tympanic and temporal with rectal measurement has shown good correlation with all methods and the correlation was highest for temporal artery measurement in this study.16-20
A comparison of temperatures measured by the commonly used methods like digital axillary and infra-red temporal and aural were undertaken for their correlation. Intra aural measurement had a good correlation with infra-red temporal and axillary. All the three methods did not show any statistically significant difference in temperature recorded. There was a significant correlation between intra aural with temporal touch and axillary method. The infra-red temperature measured at 3 cm distance showed a significantly lower temperature in comparison with the axillary and the temporal touch values. A comfortable method of temperature measurement in children will be temporal artery measurement which does not need child compliance as compared to intra aural and axillary.
CONCLUSION
Among the various methods of temperature measurement infra-red temporal touch and digital axillary have good correlation with intra aural and hence can be reliably used in children. Since infra-red temporal touch is much easier technique than digital axillary, temporal measurement may be preferred in children.
Funding: No funding sources
Conflict of interest: None declared
Ethical approval: The study was approved by the Institutional Ethics Committee
REFERENCES
- 1. Davis T. NICE guideline: Feverish illness in children-assessment and initial management in children younger than 5 years. Arch Dis Child Educ Pract Ed. 2013;98:232-5.
- 2. El-Radhi AS, Barry W. Thermometry in pediatric practice. Arch Dis Child. 2006;91:351-6.
- 3. Dew P.L. Is tympanic membrane thermometry the best method for recording temperature in children? J Child Health Care. 2006;10:96-110.
- 4. Fraden J, Lackey RP. Estimation of body sites temperatures from tympanic measurements. Clin Pediatric. 1991;30(suppl):65-70.
- 5. García Callejo FJ, Platero Zamarreño A, Sebastián Gil E, Marco Sanz M, Alpera Lacruz RJ, Martínez Beneyto MP. Otologic determining factors on infra-red tympanic thermometry in children. Acta Otorrinolaringol Esp. 2004;55:107-113.
- 6. Sehgal A, Jyothi MC, Dubey NK. Comparison of tympanic and rectal temperatures in febrile children. Indian Pediatr. 2003;40:135-40.
- 7. Pickersgill J, Fowler H, Boothman J, Thompson K, Wilcock S, Tanner J. Temperature taking: Children's preferences. Pediatr Nurs. 2003;15:22-25.
- 8. Bahorski J., Repasky T., Ranner D. Temperature measurement in pediatrics: a comparison of the rectal method versus the temporal artery method. J Pediatr Nurs. 2012;27:243-7.
- 9. Paes B.F., Vermuelen K., Brohet R.M. Accuracy of tympanic and infrared skin thermometers in children. Arch Dis Child. 2010;95:974-8.
- 10. Haugan B1, Langerud AK, Kalvoy H, Froslie KF, Riise E, Kapstad H Can we trust the new generation of infrared tympanic thermometers in clinical practice? J Clin Nurs. 2013;22(5-6):698-709.
- 11. García Callejo FJ, Platero Zamarreno A, Sebastian Gil E, Marco Sanz M, Alpera Lacruz RJ, Martínez Beneyto MP. Otologic determining factors on infra-red tympanic thermometry in children. Acta Otorrinolaringol Esp. 2004;55:107–113.
- 12. Edelu BO, Ojinnaka NC, Ikefuna AN. A comparison of axillary with rectal thermometry in under 5 children. Niger Med J. 2011;52(4):207-10.
- 13. Lodha R, Mukerji N, Sinha N, Pandey RM, Jain Y. Is axillary temperature an appropriate surrogate for core temperature? Indian J Pediatr. 2000;67:571-4.
- 14. Chaturvedi D, Vilhekar KY, Charturvedi P, Bharambe MS. Comparison of axillary temperature with rectal or oral temperature and determination of optimum placing time in children. Indian Pediatr. 2004;41:600-3.
- 15. Asadi-Pooya A, Kashef S. Accuracy of axillary temperature compared with rectal temperature in sick children. Kuwait Med J. 2006;38:25-7.
- 16. Batra P, Goyal S. Comparison of rectal, axillary, tympanic, and temporal artery thermometry in the pediatric emergency room. Pediatric emergency care. 2013;29(1):63-6.
- 17. Nimah MM1, Bshesh K, Callahan JD, Jacobs BR. Infrared tympanic thermometry in comparison with other temperature measurement techniques in febrile children. Pediatr Crit Care Med. 2006;7(1):48-55.
- 18. Reynolds M, Bonham L, Gueck M, Hammond K, Lowery J, Redel C, et al. Are temporal artery temperatures accurate enough to replace rectal temperature measurement in pediatric ED patients? J Emerg Nurs. 2014; 40(1):46-50.
- 19. Bahorski J, Repasky T, Ranner D, Fields A, Jackson M, Moultry L, et al. Temperature measurement in pediatrics: a comparison of the rectal method versus the temporal artery method. J Pediatr Nurs. 2012; 27(3):243-7.
- 20. Teran CG, Torrez-Llanos J, Teran-Miranda TE, Balderrama C, Shah NS, Villarroel P. Clinical accuracy of a non-contact infrared skin thermometer in pediatric practice.Child Care Health Dev. 2012;38(4):471-6.