Original Research ArticleInternational Journal of Community Medicine and Public Health

Publication:
Utilization of sinogram affirmed iterative reconstruction on 128 multi slice computed tomography scan to reduce radiation dose and improve image quality on thorax multi slice computed tomography scan: chest phantom study

Int J Community Med Public Health. 2019;6(10):4533-4537.DOI10.18203/2394-6040.ijcmph20194525

Thumbnail Image
View on the journal's website(opens in a new tab)

Abstract

Background Thorax MSCT examination is a diagnostic imaging that is capable of displaying both normal and pathological lung and respiratory organs. MSCT examination also has a better level of sensitivity and specificity compared to other modalities, but the radiation exposure given is very high, so the radiation dose given to patients is high. The reduction in radiation dose is very important because of the direct exposure to sensitive tissue. One method of reducing radiation dose is by reducing the tube voltage. However, the decrease in tube voltage causes a decrease in image quality as indicated by increased noise and decreased CNR. To maintain the quality of the image at low tube voltage setting, an IR reconstruction of SAFIRE was used. The purpose of this research is to know the impact of using SAFIRE on dose radiation and image quality of thorax MSCT.

Methods This study was an experimental study with a quasi-experimental study design. The object used was the N-1 Lungman chest phantom in which an artificial tumor was attached. Radiation dose assessment used CTDI value, while image quality assessment used noise and CNR. Data processing was conducted using linear regression test.

Results There was an effect of tube voltage setting and SAFIRE setting on radiation dose and image quality.

Conclusions Tube voltage ssetting and SAFIRE setting had an effect on radiation dose and image quality. Tube voltage setting and SAFIRE strength level setting that were able to provide optimal radiation dose and image quality were tube voltage of 80 kVp and SAFIRE strength levels 3 and 4 (S3 and S4).

Keywords

Affiliations

  1. Department of Imaging Diagnostic, Health Polytechnic Semarang, Semarang, Indonesia
  2. Department of Physics, Diponegoro University, Semarang, Indonesia

Full item record