Case ReportInternational Surgery Journal

Publication:
Spontaneous uncoupling of microvascular anastomotic coupler device causing post-operative haematoma - salvage with re-coupling of the device: a case report

Int Surg J. 2025;12(3):399-402.DOI10.18203/2349-2902.isj20250576

Thumbnail Image
View on the journal's website(opens in a new tab)

Abstract

We report here a case of spontaneous uncoupling of the microvascular anastomotic coupler device (MACD), commonly known as venous coupler, causing acute post-operative bleeding and haematoma in early post-operative period, in a patient with free muscle sparing TRAM flap. This was noticed as the patient suddenly abducted her right arm post-operatively while patient tried to get up in a bed. This acute emergency was salvaged by re-engaging the coupler rings, with a simple technique with reinforced 9-0 Nylon sutures, without the need for revising the entire anastomosis.

Keywords

Affiliations

  1. Department of Plastic Surgery, Gold Coast University Hospital, Gold Coast, Queensland, Australia

Full item record