Original Research ArticleInternational Journal of Community Medicine and Public Health

Publication:
A comparative study of ProRithm and standard monitoring techniques for non-invasive blood pressure measurement using photoplethysmography and electrocardiography signals through artificial intelligence/machine learning methods

Int J Community Med Public Health. 2024;11(7):2637-2641.DOI10.18203/2394-6040.ijcmph20241611

Thumbnail Image
View on the journal's website(opens in a new tab)

Abstract

Background Multi-parameter monitoring devices are essential for providing real-time patient data, which is crucial for effective healthcare interventions. This clinical trial evaluated the accuracy of the ProRithm beat-to-beat cuffless device for arterial blood pressure monitoring, comparing it with a standard sphygmomanometer.

Methods This observational study included 30 subjects aged 18 and above. Systolic and diastolic blood pressure measurements from both the ProRithm device and the Philips Monitor were compared using statistical analysis.

Results The analysis revealed no statistically significant differences between the ProRithm device and the manual method. In comparison with manual measurements using a sphygmomanometer, the mean systolic blood pressure was 131.2 mmHg with ProRithm it was 129.3 mmHg. Similarly, with the manual method, while the mean diastolic blood pressure was 76.2 mmHg and with ProRithm it was 75.9 mmHg.

Conclusions This study indicates that portable, small-sized devices like ProRithm, which facilitate remote monitoring, are effective for real-time blood pressure assessment in clinical settings.