International Journal of Otorhinolaryngology and Head and Neck Surgery

Senniappan S et al. Int J Otorhinolaryngol Head Neck Surg. 2020 Oct;6(10):1878-1881

http://www.ijorl.com

pISSN 2454-5929 | eISSN 2454-5937

Utility of Rinne’s tuning fork test for quantitative assessment of conductive hearing loss

Shivakumar Senniappan*, Rohith Chendigi


Department of Otorhinolaryngology, Vinayaka Missions Kirupananda Variyar Medical College and Hospitals, Salem, Tamil Nadu, India

Received: 20 August 2020

Accepted: 05 September 2020

*Correspondence:

Dr. Shivakumar Senniappan,
E-mail: drshiva73@gmail.com

Copyright: © the author(s), publisher and licensee Medip Academy. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.

ABSTRACT

Background: The Rinne tuning fork test is used routinely in clinical ENT examination. It is used to assess the person’s hearing acuity. Ideally, 3 tuning forks are used 256, 512 and 1024 Hz. Rinne tuning fork tests can be used to diagnose conductive hearing loss. Aim of our study an attempt is made to find the utility of three different tuning forks (256,512,1024 Hz) for quantification of conductive hearing loss and their accuracy.

Methods: A retrospective study with a sample size of 300 was undertaken over 12 months from April 2018 to March 2019 at the out-patient department of ENT of VMKV medical college and hospital, Salem. Patients of both sexes and aged above 10 yrs presenting with conductive hearing loss due to varied etiology were subjected to complete ENT examination including Rinne test with three different tuning forks (256, 512, 1024 Hz).

Results: The results of all the tuning forks (256, 512, 1024 Hz) shows that 29 patients were Rinne’s positive to all tuning forks had the air-bone gap range of 15 dB to 19 dB with a mean air-bone gap of 17.63 dB. 83 patients were negative to 256 Hz tuning fork but positive to 512 and 1024 Hz tuning forks had the air-bone gap range of 20 to 29 dB with a mean air-bone gap of 25.46 dB.

Conclusions: We conclude that Rinne’s tuning fork test can be used to quantify the degree of conductive hearing loss into mild (20-30 dB), moderate (30-45 dB), and severe (45-60 dB).

Keywords: Rinne test, Tuning forks, Conductive hearing loss, Utility

INTRODUCTION

The standard loudness comparison method of performing the Rinne test compares the relative loudness of air conduction against bone conduction. In ears with normal hearing or pure sensorineural hearing impairment, the tuning fork will sound louder opposite the ear canal (air conduction) than when placed on the mastoid bone behind the ear (bone conduction). Therefore, with the tuning fork activated, the patient is asked to say which of the following sounds louder: the fork placed within 2cm of the external auditory canal or placed firmly on the flat bone of the mastoid behind the pinna.1 To aid bone contact with the tuning fork base it is advisable to hold the patient’s head with the other hand to prevent it from moving. The alternative method of performing the Rinne test is the threshold comparison method. Here, the activated fork is held opposite the external canal until it is no longer heard. It is then placed on the mastoid bone.2 If the sound is heard again, then bone conduction is better than air conduction and there is a conductive impairment. This method of performing the Rinne is not as reliable as the relative loudness comparison method. For the Rinne test, a vibrating tuning fork (typically 512 Hz) is placed initially on the mastoid process behind each ear until the sound is no longer heard. Then, without re-striking the fork, the fork is then quickly placed just outside the ear with the patient asked to report when the sound caused by the vibration is no longer heard.3 A normal or positive Rinne test is when sound is still heard when the tuning fork is moved to air near the ear (air conduction or AC), indicating that AC is equal or greater than (bone conduction or BC). Therefore, AC>BC; which is how it is reported clinically for a normal or positive Rinne result. In the conductive hearing loss, bone conduction is better than air or BC>AC, a negative Rinne, and the patient will report that they do not hear the fork once it is moved.4 The Rinne test is not ideal for distinguishing sensorineural hearing loss, as both sensorineural hearing loss and normal hearing report a positive Rinne test (though the sensorineural patient will have a decreased duration of hearing sound once the fork is moved to air).Although many clinicians use a 516 Hz tuning fork, the 256 Hz fork gives superior sensitivities and specificities.5 The 256 Hz Rinne tuning fork test will detect a conductive defect above 30 dB in 90% of patients. Between 20 and 30 dB, the sensitivity will fall to 70% and between 10 and 20 dB, it will be less than 50%. The specificity of the test is high above 30 dB conductive defect but falls as the air-bone gap narrows. By using 256 and 1024 Hz tuning fork along with 512 Hz tuning fork conductive hearing loss can be quantified. This quantification value varies, as given by Miltenburg.6

METHODS

A retrospective study with a sample size of 300 patients was done over 12 months from April 2018 to March 2019 at OPD of ENT department of VMKVMCH, Salem. Patients of both sexes and aged above 10 years and well cooperative with complaints of conductive hearing loss due to varied etiology were included in our study. All patients were subjected to complete ENT clinical examination, routine blood investigations. Following this Rinne’s tuning fork test was done to all patients with three tuning forks (256, 512, 1024 Hz). The results were obtained, calculated, compared and correlated with an audiometric air-bone gap in PTA and terms of mild, moderate and severe conductive hearing loss

Statistical analysis

Data entry was made in the Microsoft Excel software in codes and analysis was done with an SPSS-20 computer package. Categorical variables are expressed as percentages whereas continuous variables are expressed as mean ± standard deviation. Association between the categorical variable was found by the chi-square test and the relationship between the continuous variable was assessed by Student’s t-test p<0.05 was considered as statistically significant.

RESULTS

It is seen from (Table 1) that the majority of the population is in the age group of 41-50, which shows that the majority of the population are males and the male to female ratio is 1.2:1. The mean age of the study population was 44.5 years.

Table 1: Age wise distribution of study population.
Table reproduced from source PDF page 2

The frequency of causes for conductive hearing loss (Table 2), which shows CSOM as the most frequent cause for conductive hearing loss. Adhesive otitis media is the second most frequent cause followed by traumatic perforation followed by impacted wax followed by otosclerosis. Tympanosclerosisis the least cause of conductive hearing loss.

Table 2: Disease distribution among the study population.
Table reproduced from source PDF page 2

Tuning fork test results (Table 3) show that among 300 patients 270 were negative to 256 Hz tuning fork, 189 were negative to 512 Hz and 63 were negative to 1024. Majority of patients were negative to 256 Hz.

Table 3: Tuning fork test results among study group.
Table reproduced from source PDF page 2

The results of all the tuning forks (256, 512, 1024 Hz) as shown in Table 4 which shows that 29 patients were Rinne’s positive to all tuning forks had the air-bone gap range of 15 dB to 19 dB with a mean air-bone gap of 17.63 dB. 83 patients were negative to 256 Hz tuning fork but positive to 512 and 1024 Hz tuning forks had the air-bone gap range of 20 to 29 dB with a mean air-bone gap of 25.46 dB. 125 patients were negative to 256, 512 Hz tuning forks but negative to 1024 Hz tuning fork with air-bone gap range of 31 to 44 dB with a mean air-bone gap of 37.22 dB. 63 patients were negative to all the three tuning forks with air-bone gap range of 46-59 dB mean air-bone gap of 49.73 dB and the p value is 0.0001 (p<0.005).

Table 4: Combined results of Rinne’s tuning fork test with 256, 512, 1034 Hz tuning forks.
Table reproduced from source PDF page 3
Table 5: Range of hearing loss corresponding to degree of hearing loss.
Table reproduced from source PDF page 3
Table 6: Comparison of degree of conductive hearing loss of both pure tone audiometry and Rinne tuning fork test.
Table reproduced from source PDF page 3

The patients were subjected to pure tone audiometry and the frequency distribution of conductive hearing loss is shown in Table 5, which shows that majority of patients have moderate conductive hearing loss (42.3%) followed by mild conductive hearing loss (29%) followed by severe conductive hearing loss (22.3%). The mean air-bone gap in pure tone audiometry was 35.36.

Comparison of the results of the degree of conductive hearing loss recorded with pure tone audiometry is compared in Table 6. It shows that the results obtained with Rinne tuning fork test are same as that of pure tone audiometry i.e. 67 patients had mild conductive hearing loss with both Rinne tuning fork test and pure tone audiometry, 104 patients had moderate conductive hearing loss with both Rinne tuning fork test and pure tone audiometry and 51 patients had severe conductive hearing loss with both Rinne tuning fork test and pure tone audiometry. The p value was 0.0001 (<0.005).

DISCUSSION

The purpose of the outer ear is to direct sounds onto the tympanic membrane. The sound vibrations are then transmitted through the middle ear via the ossicular chain before it reaches the cochlea. The cochlea plays an important role in transducing these vibrations into nerve impulses via the auditory nerve (vestibulocochlear nerve) which is then delivered along the central pathways to the auditory cortex where it is processed and perceived as sound. This pathway is termed air conduction. However, sound can also be transmitted via bone conduction where vibrations are transmitted via the skull and delivered directly to the cochlea which is buried within the temporal bone.7 Hearing loss may occur due to interruption at any point along these pathways. Rinne’s tuning fork test is the most commonly used test in OPD to identify conductive hearing loss. The results of Rinne’s tuning fork test can be confirmed with pure tone audiometry. In our study, we compared the results of the degree of conductive hearing loss obtained with the Rinne tuning fork test with that of results obtained with pure tone audiometry. In our study majority of patients were males (55.0%) and females were (45%). The most frequent age group was 41-50 years (40.7%) and the mean age was 44.5.8 Kelly et al found in their study that CSOM is the common cause of adult conductive hearing loss. Our study is on par with it and CSOM at a frequency of 72.3% (217 patients). When patients were subjected to Rinne’s tuning fork test, 270 patients (90%) were negative for 256 Hz tuning fork, 189 patients (63%) were negative for 512 Hz tuning fork and 63 patients (21%) were negative to 1024 Hz tuning fork.9 The results of all the tuning forks (256, 512, 1024 Hz) show that 29 patients were Rinne’s positive to all tuning forks had the air-bone gap range of 15 dB to 19 dB with a mean air-bone gap of 17.63 dB.10 Eighty three patients were negative to 256 Hz tuning fork but positive to 512 and 1024 Hz tuning forks had the air-bone gap range of 20 to 29 dB with a mean air-bone gap of 25.46 dB. 125 patients were negative to 256, 512 Hz tuning forks but negative to 1024hz tuning fork with air-bone gap range of 31 to 44db with a mean air-bone gap of 37.22 dB.11 Sixty three patients were negative to all the three tuning forks with the air-bone gap range of 46-59 dB mean air-bone gap of 49.73 dB and the p value is 0.0001 (p<0.005). this is in par with the values given by Miltenburg.12 When compared the values of Rinne tuning fork test with that of pure tone audiometry values, we found that patients with mild conductive hearing loss were 67, patients with moderate conductive hearing loss were 104, and patients with severe conductive hearing loss were 51 and the p value is 0.0001 (<0.005). this is in par with Browning, Swan IRC6 study on sensitivity and specificity of 256 Hz and 512 Hz tuning fork showing that our study was significant and Rinne’s tuning fork test can be used to quantify the degree of conductive hearing loss.13-15

CONCLUSION

We conclude in our study that Rinne’s tuning fork test can be used to quantify the degree of conductive hearing loss into mild, moderate, and severe when performed with 256 Hz, 512 Hz and 1024 Hz tuning forks. This will help give the clinician an idea about the degree of hearing loss before the patient is subjected to pure tone audiometry and also useful in places where the facility of pure tone audiometry is not available. This will guide the clinician to the need for further examination, investigation, and management. Also, the Weber and Rinne tuning fork tests can be used to confirm audiometric findings, particularly when the audiogram is not consistent with clinical findings. In the assessment of a patient with bilateral conductive hearing loss, the Weber test is a quick and useful test for the otorhinolaryngology (ENT) surgeon to help determine which side of the ear to operate on first. Usually, the ear with the more significant conductive hearing loss is preferred.

Funding: No funding sources

Conflict of interest: None declared

Ethical approval: The study was approved by the Institutional Ethics Committee

REFERENCES

  1. 1. Aarhus L, Tambs K, Kvestad E, Engdahl B. Childhood otitis media: a cohort study with 30-year follow-up of hearing (the hunt study). Ear and Hearing. 2017;36(3):302-8.
  2. 2. Browning GG, Swan IRC, Chew KK. Clinical role of informal tests of hearing. Journal of Laryngology & Otology. 2008;103:7-11.
  3. 3. Browning GG, Swan IRC. Sensitivity & specificity of the Rinne tuning fork test. British Medical Journal. 2018;297:1381-2.
  4. 4. Burkey JM, Lippy WH, Schuring AG, Rizer FM. Clinical utility of the 512-Hz Rinne tuning fork test. Am J Otol. 2008;19(1):59-62.
  5. 5. Chole RA, Cook GB. The Rinne test for conductive deafness: a critical reappraisal. Arch Otolaryngol Head Neck Surg. 2008;114(4):399-403.
  6. 6. Doyle PJ, Anderson DW, Sipke P. The tuning fork: An essential instrument in otologicpractice. J Otolaryngol. 2009;13:83-6.
  7. 7. Hardyal SRS, Hashim HZ, Mohamad I. A rare complication of tuning fork test. Indian J Otol. 2017;23:264-6.
  8. 8. Huizing EH. The early descriptions of the so-called tuning fork tests of weber, Rinne, Schwabach, and Bing II. The “Rhine test” and its first description by Polansky. J Otorhinolaryngol Relat Spec. 2009;37:88-91.
  9. 9. Kelly EA, Li B, Adams ME. Diagnostic Accuracy of Tuning Fork Tests for Hearing Loss: A Systematic Review. Otolaryngol Head Neck Surg. 2018;159(2):220-30.
  10. 10. Miltenburg DM. The validity of tuning fork tests in diagnosing hearing loss. J Otolaryngol. 2014;23(4):254-9.
  11. 11. Pearce JM. Early days of the tuning fork. J NeurolNeurosurg Psychiatry. 2011;65:728-33.
  12. 12. Rabinowitz PM. Noise-induced hearing loss. Am Fam Physician. 2000;61:58-60.
  13. 13. Rinne AB, ZurPhysiologie des menschlichen others: Bierteljahrschrift fur die praktischeHeilkundeInHalla J Hasner JV, eds. Herausgeben Von Der Medicinischen is facultative in PragVol 45. Progue: Kark Andre Publishers; 2005;20:4.
  14. 14. Meckechnie CA, Greenberg JJ, Mccall AA, Hirsch BE, Durrant JD. Rinne revisited: steel versus aluminum tuning forks. Razy Otolaryngol Head Neck Surgery. 2013;149(6):907-13.
  15. 15. Stankiewicz JA, Mowry HJ. Clinical accuracy of tuning fork tests. Laryngoscope. 2009;89(12):1956-63.