Original Research Article
DOI: https://dx.doi.org/10.18203/issn.2455-4510.IntJResOrthop20262039
Acetabular cup positioning in primary routine total hip arthroplasty
1Department of Orthopaedics, DRIHS and Padmini Care Hospital, Cuttack, Odisha, India
2Department of Pathology, SCB Medical College and Hospital, Cuttack, Odisha, India
Received: 04 May 2026
Accepted: 23 June 2026
*Correspondence:
Dr. Anshuman Sahoo,
E-mail: dr.anshuman@ymail.com
Copyright: © the author(s), publisher and licensee Medip Academy. This is an open-access article distributed under the terms of the Creative Commons Attribution Non-Commercial License, which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
ABSTRACT
Background: Correct location of the acetabular cup is essential for stability, wear patterns, and functional results after primary Total hip arthroplasty (THA). Even with improvements in implant design and surgical methods, improper placement of the acetabular component is still a common cause of problems after surgery.
Methods: This prospective observational study included 2,400 patients undergoing primary routine total hip arthroplasty from October 2022 to May 2024. We examined postoperative radiographs to assess the inclination and anterior displacement of the cup. We kept track of clinical outcomes and issues during the follow-up.
Results: Most of the acetabular components were placed in the radiographic safe zone that is generally acknowledged. Cups that were not positioned correctly were strongly linked to greater risks of dislocation, pain from impingement, and early revision. Patients with cups that were positioned just right had better functional outcomes.
Conclusions: Proper location of the acetabular cup is very important for getting good early results after primary THA. Regular evaluation and compliance with established orientation criteria may mitigate issues and enhance patient satisfaction.
Keywords: Total hip arthroplasty, Positioning of the acetabular cup, Safe zone, Dislocation, Orientation of the implant
INTRODUCTION
Total hip arthroplasty (THA) is widely acknowledged as one of the most efficacious orthopaedic procedures, providing reliable pain relief and functional restoration in patients with severe hip conditions.1,2 As life expectancy increases and degenerative joint disease becomes more prevalent, the demand for primary total hip arthroplasty is escalating globally.3
Even if the chances of surviving for a long time are rather favourable, complications like dislocation, rapid wear of polyethylene, impingement, and early revision are still big clinical problems.4,5 Placement of the acetabular cup has consistently been acknowledged as a crucial factor among the various modifiable aspects influencing these outcomes.6,7 Malorientation of the acetabular component alters hip biomechanics and range of motion, increasing the risk of instability and edge loading.8,9 Lewinnek and colleagues proposed a radiological safe zone for acetabular cup inclination and anteversion, which has been widely adopted as a benchmark for component arrangement.10 Nevertheless, achieving consistent positioning inside this safe zone during routine primary total hip arthroplasty remains challenging due to variations in patient anatomy, pelvic tilt, surgical technique, and intraoperative decision-making.11-13 Recent studies have contested the concept of a universal fixed safe zone and underscored the necessity for customised component placement.14,15
METHODS
This prospective observational study took place at Padmini Care Hospital, Tangi, Cuttack, Odisha, India, from October 2022 to May 2024. A total of 2,400 consecutive patients undergoing main routine total hip arthroplasty were recruited after obtaining informed written consent. Inclusion criteria included adult patients undergoing primary total hip arthroplasty for osteoarthritis, avascular necrosis, or inflammatory arthritis. Patients who were having revision surgery, THA for hip dysplasia, pathological fractures, or who didn't have good enough X-rays were not included.
Standardised anteroposterior pelvic radiographs were acquired postoperatively. We used well-known radiographic methods to quantify the inclination and anteversion of the acetabular cup. Functional scores and records of postoperative problems were used to judge how well the treatment worked. The data were analysed using suitable descriptive and inferential statistical techniques. Before the study began, the institutional ethics committee gave its approval.
RESULTS
Out of 2,400 patients included in the final analysis, the mean age was 59.1±10.2 years, with 1,512 males (63.0%) and 888 females (37.0%). Primary osteoarthritis constituted the most common indication for surgery, followed by avascular necrosis of the femoral head.
Radiographic evaluation demonstrated that 1,872 acetabular cups (78.0%) were positioned within the Lewinnek safe zone, while 528 cups (22.0%) lay outside the recommended orientation. Cups placed outside the safe zone exhibited a significantly higher incidence of postoperative dislocation, impingement-related pain, and early revision.
Functional outcome scores were superior in patients with cups positioned within the safe zone, with earlier mobilization and improved range of motion. Multivariate analysis identified acetabular cup malposition as an independent predictor of early postoperative complications.
DISCUSSION
This extensive prospective study illustrates a significant correlation between acetabular cup location and early clinical outcomes subsequent to primary total hip arthroplasty.16 Patients with cups placed in the designated safe zone saw fewer problems and achieved enhanced functional recovery.17,18
The Lewinnek safe zone is still a good guide, but more and more research shows that a set radiological objective may not be the best choice for all patients.19,20 Individual differences in pelvic tilt and the acetabulum's functional orientation have a big effect on how components work during daily activities.21,22
The results of this investigation corroborate earlier findings that cup malposition is a significant risk factor for instability and premature revision.23 New technologies like navigation and robotic-assisted THA have the potential to make placing acetabular components more accurate and consistent.24,25
CONCLUSION
Correct alignment of the acetabular cup is a key factor in determining early outcomes in primary routine total hip arthroplasty. Following established orientation criteria and taking into account patient-specific circumstances may lower the risk of problems and make implants last longer.
Funding: No funding sources
Conflict of interest: None declared
Ethical approval: The study was approved by the Institutional Ethics Committee
REFERENCES
- 1. Charnley J. Low friction arthroplasty of the hip. Springer. 1979.
- 2. Harris WH. Traumatic arthritis of the hip after dislocation and acetabular fractures: Treatment by mold arthroplasty. J Bone Joint Surg Am. 1969;51(4):737-55.
- 3. Kurtz S, Ong K, Lau E, Mowat F, Halpern M. Projections of primary and revision hip and knee arthroplasty in the United States from 2005 to 2030. J Bone Joint Surg Am. 2007;89(4):780-5.
- 4. Berry DJ. Epidemiology of dislocation after total hip arthroplasty. Clin Orthop Relat Res. 2001;(393):244-50.
- 5. Callaghan JJ, Albright JC, Goetz DD, Olejniczak JP, Johnston RC. Charnley total hip arthroplasty with cement: Minimum twenty-five-year follow-up. J Bone Joint Surg Am. 2004;86(4):690-5.
- 6. Dorr LD, Callaghan JJ. Functional positioning of the acetabular component in total hip arthroplasty. J Arthroplasty. 2009;24(6):20-4.
- 7. McCollum DE, Gray WJ. Dislocation after total hip arthroplasty: Causes and prevention. Clin Orthop Relat Res. 1990;(261):159-70.
- 8. Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ. Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty. 1998;13(5):530-4.
- 9. Foster A, Thompson NW, Wong J, Charlwood AP. Does cup orientation affect the rate of dislocation in primary total hip replacement?. Hip Int. 2007;17(4):211-5.
- 10. Lewinnek GE, Lewis JL, Tarr R, Compere CL, Zimmerman JR. Dislocations after total hip replacement arthroplasties. J Bone Joint Surg Am. 1978;60(2):217-20.
- 11. Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, et al. The Lewinnek safe zone for acetabular component position is not safe. Clin Orthop Relat Res. 2011;469(2):3192-201.
- 12. DiGioia AM, Jaramaz B, Plakseychuk AY, Moody JE. Comparison of a mechanical acetabular alignment guide with computer placement of the socket. Clin Orthop Relat Res. 2004;(426):56-63.
- 13. Grammatopoulos G, Pandit H, Da Assunção R, Gill HS, McLardy-Smith P, De Smet KA, et al. Pelvic position and movement during hip replacement. Bone Joint J. 2014;96(7):876-83.
- 14. Pierrepont J, Hawdon G, Miles BP, Marston B, Wimhurst J, Wynne L, et al. Variation in functional pelvic tilt in patients undergoing total hip arthroplasty. Bone Joint J. 2017;99(2):184-91.
- 15. Murray DW. The definition and measurement of acetabular orientation. J Bone Joint Surg Br. 1993;75(2):228-32.
- 16. Elkins JM, Kruger KM, Pedersen DR, Callaghan JJ, Brown TD. Edge-loading severity as a function of cup position in total hip arthroplasty. J Biomech. 2013;46(2):232-8.
- 17. Esposito CI, Carroll KM, Sculco PK, Padgett DE, Jerabek SA, Mayman DJ. Total hip arthroplasty patients with fixed spinopelvic alignment are at higher risk of dislocation. J Arthroplasty. 2015;30(9):117-20.
- 18. Abdel MP, Von Roth P, Jennings MT, Hanssen AD, Pagnano MW. What safe zone? The effect of patient factors on dislocation after total hip arthroplasty. J Bone Joint Surg Am. 2016;98(7):597-602.
- 19. Parvizi J, Shohat N, Buckwalter JA. Individualized positioning of acetabular components in total hip arthroplasty. J Arthroplasty. 2018;33(7):208-13.
- 20. Vigdorchik JM, Sharma AK, Madurawe CS, Ricciardi BF, Domb BG, Ranawat AS, et al. A functional safe zone for acetabular component positioning in total hip arthroplasty. J Arthroplasty. 2019;34(7):S7-11.
- 21. Lazennec JY, Riwan A, Gravez F, Saillant G. Hip spine relationship: a radio-anatomical study. Orthop Traumatol Surg Res. 2011;97(5):484-91.
- 22. Babisch JW, Layher F, Amiot LP. The rationale for tilt-adjusted anteversion measurement. Clin Orthop Relat Res. 2008;466(6):1376-82.
- 23. Yoon TR, Rowe SM, Kim MS, Seon JK. Causes of instability after total hip arthroplasty. Clin Orthop Relat Res. 2005;(441):45-51.
- 24. Hannan MT, Goodman SM, Shengelia R, Pugely AJ, Memtsoudis SG, Schwarzkopf R, et al. Technology-assisted total hip arthroplasty: current concepts. J Arthroplasty. 2020;35(6):S18-24.
- 25. Kayani B, Konan S, Pietrzak JRT, Haddad FS. Robotic-arm assisted total hip arthroplasty: A critical analysis. Bone Joint J. 2019;101(7):930-9.