Original Research ArticleInternational Journal of Research in Orthopaedics
Publication: What’s new in the Visionaire system® in second-generation for TKA? a comparative study of accuracy, efficiency and functional results
Int J Res Orthop. 2022;8(3):298-302.DOI10.18203/issn.2455-4510.IntJResOrthop20221117
Abstract
Background The aim of the study was to analyze and compare the accuracy, efficiency and functional evolution between the first- and second-generation Patient-specific instrumentation (PSI).
Methods We report our experience in TKA using PSI Visionaire System® (Smith and Nephew®) in 456 procedures. The patients were divided into first- (N=272) and second-generation PSI design (N=184). For the accuracy was analyzed the mechanical results, namely Hip-knee-ankle (HKA) alignment post-TKA and outliers’ frequency; for the efficiency was analyzed the length time surgery, the length of stay and satisfaction, for the functional outcomes was compared the evolution at pain, range of motion, gait perimeter and the domains of the Western ontario and mcmaster osteoarthritis index (WOMAC). The t-test for independent samples was applied in the continuous numeric variables and Qui square tests in the dichotomic nominal variables. Statistical significance was p value 0.001) and in the third in WOMAC stiffness (p<0.001).
Conclusions The second-generation PSI of Visionaire system® improved the alignment of the tibia, but not the alignment of the femur. This change did not produce any improvements in functioning.
Keywords
Affiliations
- Department of Orthopaedics, Hospital Particular do Algarve, Faro, Portugal Health School of Algarve University, Faro, Portugal
- Department of Orthopaedics, Hospital Particular do Algarve, Faro, Portugal